XI'AN BEICHENG ELECTRONICS CO.,LTD

The Best Quality, The Best Service!

Manufacturer from China
Site Member
5 Years
Home / Products / Multifunction Process Calibrator /

VICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibrator

Contact Now
XI'AN BEICHENG ELECTRONICS CO.,LTD
Country/Region:china
Contact Person:Mrssales
Contact Now

VICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibrator

Ask Latest Price
Video Channel
Brand Name :VICTOR
Model Number :VICTOR 07
Certification :CE
Place of Origin :CHINA
MOQ :1pcs
Price :USD268/pc
Payment Terms :T/T, Western Union, L/C, Paypal
Supply Ability :3000pcs available
Delivery Time :5 to 30 days
Packaging Details :10pcs/Carton; Carton Size: 39.5*38.5*26cm; 10KG/carton
Color :Blue or Orange and Black
Size :180*90*47mm
Product Name :Loop Calibrator
Model No. :VICTOR 07
Packing Details :10pcs/Carton; Carton Size: 39.5*38.5*26cm
Accuracy :0.05% accuracy
Weight :Approx. 900 g (including batteries)
Power supply :2*1.5V AA Alkaline battery
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

VICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibrator

Introduction

Temperature calibrator (hereinafter referred to as calibrator) is a hand-held portable instrument powered by battery and capable of measuring and outputting electrical and physical parameters. Can be used for measuring DC voltage, DC current, resistance, thermocouple and thermal resistance; And simulating output DC voltage, DC current, resistance, thermocouple, thermal resistance.

Measurement function Range Measuring range Resolution Accuracy Remarks

DC Voltage

DCV

100mV -110.000mV~110.000mV 0.001mV 0.05%+0.03mV Input resistance: Approximately 1MΩ
1V -1.1000V~1.1000V 0.0001V 0.05%+0.3mV
35V -35.000V~35.000V 0.001V 0.02%+2mV

DC current

DCI

30mA -35.000mA~35.000mA 0.001mA 0.02%+4uA

Shunt resistance: Approximately 10Ω

Input resistance: Approximately 20Ω

Resistance

OHM

500Ω 0~550.00Ω 0.01Ω 0.05%+0.2Ω

2W/3W/4W measurement

500Ω~1mA excitation

5KΩ~0.1mA excitation

Open circuit voltage: Approximately 2.5V

Lead resistance is not included in measurement accuracy

5KΩ 0~5.5000KΩ 0.0001KΩ 0.05%+2Ω

Thermocouple

TC

R 0°C~1767°C 1°C

0~500℃

500~1767℃

2.5℃

2℃

Use ITS-90 temperature scale

Error of cold junction compensation is not included in accuracy

Sensor inaccuracy not included in accuracy

The effect of thermoelectric potential is not included in the accuracy

The corresponding materials of the left division number are as follows:

R: Platinum rhodium 13-Platinum

S: Platinum rhodium 10-Platinum

K: Nickel chromium-Nickel silicon

E: Nickel chromium-Copper nickel (Constantan)

J: Iron-Copper nickel (Constantan)

T: Copper-Copper-nickel (Constantan)

N: Nickel chromium silicon-Nickel silicon

B: Platinum rhodium 30-Platinum

A: Tungsten rhenium 5-Tungsten-rhenium 20

C: Tungsten rhenium 5-Tungsten-rhenium 26

D: Tungsten rhenium 3-Tungsten-rhenium 25

S 0°C~1767°C
K -100.0°~1372.0°C 0.1°C

-100.0~0.0℃

0.0~1372.0℃

1.2℃

0.8℃

E -50.0°C~1000.0°C

-50.0℃~0.0℃

0.0~1000.0℃

0.9℃

1.5℃

J -60.0°C~1200.0°C

-60.0~0.0℃

0.0~1200.0℃

1.0℃

0.7℃

T -100.0°C~400.0°C

-100.0~0.0℃

0.0~400.0℃

1.0℃

0.7℃

N -200.0°~1300.0°C

-200.0~0.0℃

0.0~1300.0℃

1.5℃

0.9℃

B 600°C~1820°C 1°C

600~800℃

800~1000℃

1000~1820℃

2.2℃

1.8℃

1.4℃

A 0°C~2500°C 1°C

0~1600℃

1600~2000℃

2000~2500℃

2.0℃

2.2℃

2.4℃

C 0°C~2310°C 1°C

0~1600℃

1600~2000℃

2000~2310℃

2.0℃

2.2℃

2.4℃

D 0°C~2310°C 1°C

0~100℃

100~270℃

270~1200℃

270~2310℃

2.6℃

2.4℃

2.2℃

2.4℃

Thermal resistance

RTD

(3W)

Pt100

385

-200.0°C~800.0°C 0.1°C

-200.0~0.0℃

0.0~400.0℃

400.0~800.0℃

0.5℃

0.7℃

0.8℃

Use Pt (385) temperature scale

2W/3W/4W measurement

500Ω~1mA excitation

5KΩ~0.1mA excitation

Open circuit voltage: Approximately 2.5V

Errors introduced by 2W/3W measurement lead resistance mismatch are not included in accuracy

Sensor inaccuracy not included in accuracy

The effect of thermoelectric potential is not included in the accuracy

Pt200

385

-200.0°C~630.0°C

-200.0~100.0℃

100.0~300.0℃

300.0~630.0℃

0.8℃

0.9℃

1.0℃

Pt500

385

-200.0°C~630.0°C

-200.0~100.0℃

100.0~300.0℃

300.0~630.0℃

0.4℃

0.5℃

0.7℃

Pt1000

385

-200.0°C~630.0°C

-200.0~100.0℃

100.0~300.0℃

300.0~630.0℃

0.3℃

0.5℃

0.7℃

Cu50 -50.0°C~150.0°C -50.0~150.0℃ 0.7℃
Continuity detection 500Ω ≤ 50Ω will sound 0.01Ω Approximately 1mA excitation

Loop power supply

LOOP

24 V 10%

VICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibratorVICTOR 07 Temperature Calibrator Multi-function process calibrator thermocouple calibrator